Characteristics
Article
—
1210
Weight, kg, not more
—
0,9
Overall dimensions in the bag, mm
—
350х140х200
Volume of inspired gas, ml, not less than
—
150
Capacity of ventilation bag, ml
—
350±100
Minute ventilation, l/min
—
5
Pressure limitation in breathing circuit, hPa
—
55±15
Resistance to inspiration, hPa
—
1
Resistance to expiration, hPa
—
2
Gas leak at pressure 30hPa, l/min, not more than
—
0,5
Все характеристики
Field of application: emergency medical services, disaster medicine services, military and extreme medicine services; anesthesiology departments, ICUs and resuscitation departments; maternity hospitals, etc.
Current use: disposable manual resuscitators are made of PVC and used for artificial pulmonary ventilation in conditions of respiratory failure of any etiology:
-for newborns КDO-МP-N, patient weight -below 4 kg.
Advantages:
-Disposable usage excludes the risk of cross contamination;
-No need for sterilization and repacking;
-Disposable masks with inflatable air cushion provide tight fit to the patient’s face.
Current use: disposable manual resuscitators are made of PVC and used for artificial pulmonary ventilation in conditions of respiratory failure of any etiology:
-for newborns КDO-МP-N, patient weight -below 4 kg.
Advantages:
-Disposable usage excludes the risk of cross contamination;
-No need for sterilization and repacking;
-Disposable masks with inflatable air cushion provide tight fit to the patient’s face.
Article
|
1210 |
Weight, kg, not more
|
0,9 |
Overall dimensions in the bag, mm
|
350х140х200 |
Volume of inspired gas, ml, not less than
|
150 |
Capacity of ventilation bag, ml
|
350±100 |
Minute ventilation, l/min
|
5 |
Pressure limitation in breathing circuit, hPa
|
55±15 |
Resistance to inspiration, hPa
|
1 |
Resistance to expiration, hPa
|
2 |
Gas leak at pressure 30hPa, l/min, not more than
|
0,5 |
Capacity of reservoir bag, ml
|
600 |